Exfo FTB-3930 MultiTest Uživatelský manuál Strana 85

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 101
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 84
Troubleshooting
Multifunction Loss Tester 79
Solving Common Problems
Unable to send a text
message.
Selected port differs
from actual port to
which fiber is
connected.
Receiving unit’s port is
in use.
Remote unit is not
compatible.
On the sending unit, set
the correct port (SM or
MM) in Setup (F
ASTEST
tab).
Wait until FASTEST
completes, turn source
and ORL meter off or exit
Probe mode (FOT-930).
Make sure remote unit is
an FOT-930, FTB-3930,
FOT-920 or FTB-3920.
On Results tab, a box
indicates “Invld” instead of
a numerical value.
The application is unable to
calculate the F
ASTEST ORL
value (e.g., at 1310 nm for
short fibers).
Perform the measurements
again.
On Results tab, a box
indicates “- - - -” instead of
a numerical value.
The calculated value makes
no sense (e.g., negative
ORL value).
Perform the measurements
again. Note that the
calculated value is usually
shown on the printed report.
When printing a report,
rightmost columns are not
printed.
The Fiber ID column is too
wide. The width of this
column is determined with
the widest name.
Rename fiber IDs that have a
long name.
When saving data in text
formats, some ORL values
are higher than the
indicated sensitivity.
Saved text files containing
symbols such as “>” may
not open properly in some
applications.
The actual measured value is
used instead, even if it is not
accurate.
Problem Possible Cause Solution
Zobrazit stránku 84
1 2 ... 80 81 82 83 84 85 86 87 88 89 90 ... 100 101

Komentáře k této Příručce

Žádné komentáře