Exfo FTB-700 for FTB-1 Uživatelský manuál Strana 177

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 329
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 176
Analyzing the Results Manually
OTDR 169
Getting Event Loss (Four-Point and Least-Square Approximation)
5. Position marker B at the beginning of the linear area following the
event to be measured.
6. Position submarker b at the end of the linear area following the event
to be measured (must not include any significant events).
Note: The loss values that are displayed depend on which calculation methods
are selected (see Selecting the Attenuation and Loss Values that Will Be
Displayed on page 161).
4
3
5
6
Event to be measured
Linear area
Linear area
Least-square
approximation
Four-point event loss
from the areas delimited
by markers
a, A, B and b
Zobrazit stránku 176
1 2 ... 172 173 174 175 176 177 178 179 180 181 182 ... 328 329

Komentáře k této Příručce

Žádné komentáře