IEEE 488.2 and Specific Command Reference
FTB-500 379
Specific Commands
:SYSTem:DATE
Description The SYSTem:DATE command is used to set the
device's internal calendar.
This is not affected by a *RST command.
Syntax :SYSTem:DATE<wsp><Year>,<Month>,<Day
>
Parameter(s)
Yea r:
The program data syntax for <Year> is defined
as a <DECIMAL NUMERIC PROGRAM DATA>
element.
The <Year> is rounded to the nearest integer. Its
range is limited by the capability of the device.
The year shall be entered as a four-digit number,
including century and millennium information.
Month:
The program data syntax for <Month> is defined
as a <DECIMAL NUMERIC PROGRAM DATA>
element.
The <Month> is rounded to the nearest integer.
Its range is 1 to 12 inclusive. The number 1
corresponds to January, 2 to February, and so on.