IEEE 488.2 and Specific Command Reference
FTB-500 385
Specific Commands
:SYSTem:TIME
Description This device has an internal clock and
implements the SYSTem:TIME command to set
the clock time over the interface.
This is not affected by a *RST command.
Syntax :SYSTem:TIME<wsp><Hour>,<Minute>,<Sec
onds>
Parameter(s)
Hour:
The program data syntax for <Hour> is defined
as a <DECIMAL NUMERIC PROGRAM DATA>
element.
The <Hour> is always rounded to the nearest
integer. It ranges from 0 to 23 inclusively. The
device accepts hour information in 24-hour
format.
Minute:
The program data syntax for <Minute> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
The <Minute> is always rounded to the nearest
integer. It ranges from 0 to 59 inclusively.