Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Uživatelský manuál Strana 447

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 659
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 446
Referencia de instrucciones SCPI
OSA 441
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:
NOISe:AUTO
Description This command controls activation of i-InBand
noise measurement for WDM analysis of the
selected channel.
At *RST, this value is not available.
At CALCulate[1..n][:WDM]:CHANnel[:DEFine],
this value is set to off (disabled).
Syntax :CALCulate[1..n][:WDM]:CHANnel:NOISe:AUTO
<wsp><Auto>
Parameter(s) Auto:
The program data syntax for <Auto> is defined
as a <Boolean Program Data> element. The
<Auto> special forms ON and OFF are accepted
on input for increased readability. ON
corresponds to 1 and OFF corresponds to 0.
The <Auto> parameter corresponds to the new
state of auto noise measurement.
0 or OFF: disable selected channel auto noise.
1 or ON: enable selected channel auto noise.