Zobrazit stránku 550
Referencia de instrucciones SCPI
OSA 545
Product-Specific Commands—Description
:INITiate:CONTinuous?
Description This query indicates if the trigger system is
continuously initiated or not.
At *RST, this value is set to off (disabled).
Syntax :INITiate:CONTinuous?
Parameter(s) None
Response Syntax <Continuous>
Response(s) Continuous:
The response data syntax for <Continuous> is
defined as a <NR1 NUMERIC RESPONSE DATA>
element.
The <Continuous> response corresponds to the
state of the trigger system continuous cycle.
0: continuous cycle is disabled.
1: continuous cycle is enabled.
Example(s) INIT:CONT ON
INIT:CONT? Returns 1 (trigger system
continuously initiated)
Notes An acquisition may still be in progress even if
INIT:CONT? returns 0. The command
STAT:OPER:BIT8:COND? shall be used to test
acquisition completion.
See Also :INITiate[:IMMediate]