Exfo FTB-7000 OTDR Series for FTB-500 Uživatelský manuál Strana 470

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 489
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 469
Referencia de instrucciones SCPI
456 OTDR
Product-Specific Commands—Description
:SOURce[1..n]:WAVelength?
Description This query returns the output wavelength of the
currently selected source, in meters.
At *RST, the wavelength that will be selected
depends on the instrument you have.
Syntax :SOURce[1..n]:WAVelength?[<wsp>MINimum|
MAXimum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MINimum|MAXimum|DEFault.
MINimum is used to retrieve the instrument's
lowest supported value.
MAXimum is used to retrieve the instrument's
highest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Wavelength>
Zobrazit stránku 469
1 2 ... 465 466 467 468 469 470 471 472 473 474 475 ... 488 489

Komentáře k této Příručce

Žádné komentáře