Exfo FTB-7000 OTDR Series for FTB-500 Uživatelský manuál Strana 410

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 493
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 409
SCPI-Befehlsreferenz
396 Serie FTB-7000
Product-Specific Commands—Description
:CONFigure[1..n]:ANAlysis:
IORefraction
Description This command sets the index of refraction that
will be used for the next acquisition.
*RST reverts this setting to default value.
Syntax :CONFigure[1..n]:ANAlysis:IORefraction<wsp>
<IOR>|MAXimum|MINimum|DEFault
Parameter(s) IOR:
The program data syntax for <IOR> is defined as
a <numeric_value> element. The <IOR>
special forms MINimum, MAXimum and DEFault
are accepted on input.
MINimum allows to set the instrument to the
lowest supported value.
MAXimum allows to set the instrument to the
highest supported value.
DEFault allows the instrument to select a value
for the <IOR> parameter.
Sets the index of refraction.
Example(s) CONF:ANA:IOR? Ex.: Returns 1.4677
CONF:ANA:IOR 1.5
CONF:ANA:IOR? Returns 1.5
Zobrazit stránku 409
1 2 ... 405 406 407 408 409 410 411 412 413 414 415 ... 492 493

Komentáře k této Příručce

Žádné komentáře