Exfo FTB-7000 OTDR Series for FTB-500 Uživatelský manuál Strana 424

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 493
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 423
SCPI-Befehlsreferenz
410 Serie FTB-7000
Product-Specific Commands—Description
:FETCh[1..n]:ASETting:RANGe?
Description This query returns the range found after an
initiate (INIT) command. Note that acquisition
mode (CONF:ACQ:MODE) must be set to
ASETting.
Since *RST clears the range value, the returned
value will be 0.
Syntax :FETCh[1..n]:ASETting:RANGe?
Parameter(s) None
Response Syntax <Range>
Response(s) Range:
The response data syntax for <Range> is
defined as a <NR3 NUMERIC RESPONSE DATA>
element.
Returns the range, in meters.
Example(s) CONF:ACQ:MODE ASET
INIT
INIT:STAT? Returns 0 when acquisition is
complete.
FETC:ASET:RANG? Ex.: Returns 1.25E+3
See Also CONFigure[1..n]:ACQuisition:RANGe?
CONFigure[1..n]:ACQuisition:PULSe?
CONFigure[1..n]:ACQuisition:DURation?
CONFigure[1..n]:ACQuisition:WAVelength:LIST?
CONFigure[1..n]:ACQuisition:RANGe:LIST?
CONFigure[1..n]:ACQuisition:PULSe:LIST?
CONFigure[1..n]:ACQuisition
Zobrazit stránku 423
1 2 ... 419 420 421 422 423 424 425 426 427 428 429 ... 492 493

Komentáře k této Příručce

Žádné komentáře