SmartR™ Features
136 MAX-635
FaultMapper
Open and the distance to Open is displayed and represented using X in
the graph.
Test Status displays the current state of measurement or an error
message.
To select the parameter values:
1. Press the up/down arrow keys to highlight the desired parameter.
2. Press the left/right arrow keys to view and select the options.
OR
3. Press on a value to open a list box of options.
4. Press to confirm the value.
Cable Setup
The Cable Setup page allows you to set the Cable
Type, Cable Length, and Cable Temperature for the
FaultMapper test. All other Cable Parameters are read
only and correspond to the current cable book entry
as selected in Setup/Cable Book/Cable Details page.
Cable Type allows you to select a cable entry
(with its associated parameters) from the current
Cable Book, to use for testing. The selection is not
saved when leaving the test. The current Cable
Book entry is shown by default.
Cable Fill read-only entry is the type of material
the cable is filled with: Aircore, Jelly, Pulp, 5 PR,
or 2 PR.
Capacitance T-R/A-B read-only entry is a value for the capacitance per
length constant.
Capacitance T/R to GND/A/B to E read-only entry is a value for the
capacitance per length to ground constant.