Remote Control (SCPI) Commands
132 PM-1600
Specific Commands
SENSe:POWer:RANGe:AUTO
Description This command sets the power measurement range to
automatic (Autorange), for the currently selected channel when
using a PM-1620. If this command is sent while an acquisition is
in progress, the “Acquisition already running” error message
will be raised.
Syntax SENS:POW:RANG:AUTO<space><boolean>
Parameters The <boolean> parameter can be
“0”- disable Autorange
“1”- enable Autorange
Example SENS:POW:RANG:AUTO 1
See also SENS:POW:RANG?, SENS:POW:RANG:AUTO?,
SENS:POW:RANG:HIGH, SENS:POW:RANG:LIST?, and
SENS:POW:RANG:LOW
Note The Autorange function can be performed in power
measurement mode only.
Remote Control (SCPI) Commands
132 PM-1600
Specific Commands
SENSe:POWer:RANGe:AUTO
Description This command sets the power measurement range to
automatic (Autorange), for the currently selected channel when
using a PM-1620. If this command is sent while an acquisition is
in progress, the “Acquisition already running” error message
will be raised.
Syntax SENS:POW:RANG:AUTO<space><boolean>
Parameters The <boolean> parameter can be
“0”- disable Autorange
“1”- enable Autorange
Example SENS:POW:RANG:AUTO 1
See also SENS:POW:RANG?, SENS:POW:RANG:AUTO?,
SENS:POW:RANG:HIGH, SENS:POW:RANG:LIST?, and
SENS:POW:RANG:LOW
Note The Autorange function can be performed in power
measurement mode only.