SCPI Commands Reference
Optical Modulation Analyzer 155
Product-Specific Commands—Description
:SENSe:DDEMod:SRATe?
Description This query returns the symbol rate (symbols per
second) for the analyzer's digital demodulator.
At *RST, this value is set to 28 Gbaud or 20
Gbaud. It’s device dependent.
Syntax :SENSe:DDEMod:SRATe?
Parameter(s) None
Response Syntax <Rate>
Response(s) Rate:
The response data syntax for <Rate> is defined
as a <NR2 NUMERIC RESPONSE DATA>
element.
Returns the symbol Rate for the analyzers digital
demodulator.
Example(s) SENSe:DDEM:SRAT 25.3
SENSe:DDEM:SRAT?
Returns: 25.3
Notes Default frequency is 10.0 GBaud.
See Also SENSe:DDEMod:AVERage:COUNt 10