Getting Started with Copper Qualification Tests
14 AXS-200/635
Regional Settings
Cable Setup
The Cable Setup tab allows you to
configure parameter values for the
cable.
Each parameter and button is
described below:
³ Cable Type defines the gauge of
the cable in use. If wire gauge is
measured in American Wire
Gauge (AWG) units, the available
choices are: AUTO, 19 AWG, 22 AWG, 24 AWG, or 26 AWG. For mm
gauge wire, the available choices are: AUTO, 0.32 mm, 0.40 mm,
0.50 mm, 0.60 mm, 0.65 mm, 0.80 mm, 0.90 mm, or 1.20 mm.
³ Cable Fill allows you to select the type of material the cable can be
filled with. Changing the selection to AIRCORE, JELLY, PULP, 5PR, or
2PR influences the cable capacitance per length, automatically
updating the Cap. T/R-GND field.
³ Capacitance T-R (nF/km or nF/mi) allows you to specify a value for the
capacitance per length constant.
³ Cap. T/R-GND (nF/km or nF/mi) allows you to specify a value for the
capacitance per length to ground constant.
³ Resistance (Ohm/km or Ohm/mi) allows you specify a value for the
resistance constant of the cable.
³ Attenuation (dB/km or dB/mi) allows you to specify a value for the
reduction in signal strength or insertion loss of the cable.
³ VOP allows you to set the velocity of propagation for the cable as a
ratio of the speed of light. Enter a value between 0.400 and 0.999.
³ Restore Defaults allows you to reset cable parameters for the selected
cable to the standard default values.