Exfo FG-750 Uživatelský manuál Strana 257

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 338
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 256
Maintenance
Fiber Guardian 243
Cleaning EUI Connectors
To clean EUI connectors:
1. Remove the EUI from the instrument to expose the connector
baseplate and ferrule.
2. Moisten a 2.5 mm cleaning tip with one drop of isopropyl alcohol
(alcohol may leave traces if used abundantly).
3. Slowly insert the cleaning tip into the EUI adapter until it comes out on
the other side (a slow clockwise rotating movement may help).
4. Gently turn the cleaning tip one full turn, then continue to turn as you
withdraw it.
5. Repeat steps 3 to 4 with a dry cleaning tip.
Note: Make sure you don’t touch the soft end of the cleaning tip.
6. Clean the ferrule in the connector port as follows:
6a. Deposit one drop of isopropyl alcohol on a lint-free wiping cloth.
3
4
5
Zobrazit stránku 256
1 2 ... 252 253 254 255 256 257 258 259 260 261 262 ... 337 338

Komentáře k této Příručce

Žádné komentáře