Справка по командам SCPI
454 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:
CENTer[:WAVelength]
Description This command sets the nominal center
wavelength of the selected WDM analysis
channel.
At *RST, this value is not available.
Syntax :CALCulate[1..n][:WDM]:CHANnel:CENTer[:WAV
elength]<wsp><Center[<wsp>M]>|MAXimum|MIN
imum|DEFault
Parameter(s) Center:
The program data syntax for <Center> is defined
as a <numeric_value> element followed by an
optional <SUFFIX PROGRAM DATA> element.
The allowed <SUFFIX PROGRAM DATA>
element is M. The <Center> special forms
MINimum, MAXimum and DEFault are accepted
on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.