Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Uživatelský manuál Strana 558

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 688
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 557
Справка по командам SCPI
552 FTB-5240S/S-P/BP
Product-Specific Commands—Description
Response(s) Resolution:
The response data syntax for <Resolution> is
defined as a <NR3 NUMERIC RESPONSE
DATA> element.
The <Resolution> response corresponds to either
the current or the MINimum/MAXimum resolution
bandwidth value for OSNR calculation.
Example(s) CALC:WDM:OSNR:BAND:RES:AUTO OFF
CALC:WDM:OSNR:BAND:RES 0.065 NM
CALC:WDM:OSNR:BAND:RES? Returns
6.500000E-011
See Also :CALCulate[1..n][:WDM]:OSNR:BWIDth|BANDwi
dth[:RESolution]:AUTO
:CALCulate[1..n][:WDM]:OSNR:BWIDth|BANDwi
dth[:RESolution]
:CALCulate[1..n][:WDM]:OSNR:
BANDwidth|BWIDth[:RESolution]?
Zobrazit stránku 557
1 2 ... 553 554 555 556 557 558 559 560 561 562 563 ... 687 688

Komentáře k této Příručce

Žádné komentáře