Справка по командам SCPI
588 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:SENSe[1..n]:AVERage:STATe
Description This command controls activation of acquired
trace averaging.
At *RST, averaging is set to off (disabled).
Syntax :SENSe[1..n]:AVERage:STATe<wsp><State>
Parameter(s) State:
The program data syntax for <State> is defined
as a <Boolean Program Data> element. The
<State> special forms ON and OFF are accepted
on input for increased readability. ON
corresponds to 1 and OFF corresponds to 0.
The <State> parameter corresponds to the new
state of trace averaging.
0 or OFF: disable averaging.
1 or ON: enable averaging.
Example(s) SENS:AVER:STAT OFF
SENS:AVER:STAT? Returns 0 (averaging is
disabled)
Notes Trace averaging is not supported by the trigger
system when continuously initiated (INIT:CONT
ON).
See Also :INITiate[:IMMediate]
:INITiate:CONTinuous
:SENSe[1..n]:AVERage:COUNt
:SENSe[1..n]:AVERage:TYPE
:SENSe[1..n]:AVERage:STATe?