Exfo FTB -5240S/5240BP Optical Spectrum Analyzer Uživatelský manuál Strana 443

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 638
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 442
SCPI Command Reference
OSA 437
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:
NOISe:WIDTh[:WAVelength]
Description This command sets the wavelength width of the
noise measurement region of the selected WDM
analysis channel.
At *RST, this value is not available.
At CALCulate[1..n][:WDM]:CHANnel[:DEFine],
width of the noise measurement region is set to
100.0 GHz.
Syntax :CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDT
h[:WAVelength]<wsp><Width[<wsp>M]>|M
AXimum|MINimum|DEFault
Parameter(s) Width:
The program data syntax for <Width> is defined
as a <numeric_value> element followed by an
optional <SUFFIX PROGRAM DATA> element.
The allowed <SUFFIX PROGRAM DATA>
element is M. The <Width> special forms
MINimum, MAXimum and DEFault are accepted
on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
Zobrazit stránku 442
1 2 ... 438 439 440 441 442 443 444 445 446 447 448 ... 637 638