SCPI Command Reference
OSA 447
Product-Specific Commands—Description
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
The <Select> parameter corresponds to a valid
channel setup index to select. The channel index
cannot be zero.
The CALCulate[1..n][:WDM]:CHANnel:COUNt?
query can be used to determine valid index
range.
Example(s) CALC:WDM:CHAN:DEL:ALL
CALC:WDM:CHAN:DEF "C_001",1525.000 NM
CALC:WDM:CHAN:NSEL 1
CALC:WDM:CHAN:SEL? Returns "C_001"
See Also :CALCulate[1..n][:WDM]:CHANnel:COUNt?
:CALCulate[1..n][:WDM]:CHANnel[:DEFine]
:CALCulate[1..n][:WDM]:CHANnel:NSELect?
:CALCulate[1..n][:WDM]:CHANnel:SELect
:CALCulate[1..n][:WDM]:DATA:CHANnel:NSELec
t
:CALCulate[1..n][:WDM]:CHANnel:
NSELect