Exfo FTB -5240S/5240BP Optical Spectrum Analyzer Uživatelský manuál Strana 508

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 638
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 507
SCPI Command Reference
502 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:DATA:TPOWer?
Description This query returns computed WDM analysis
global result for analyzed trace total power.
At *RST, this value is not available.
Syntax :CALCulate[1..n][:WDM]:DATA:TPOWer?
Parameter(s) None
Response Syntax <Power>
Response(s) Power:
The response data syntax for <Power> is
defined as a <NR3 NUMERIC RESPONSE DATA>
element.
The <Power> response corresponds to the
computed total power of the trace.
Example(s) <Do measurement>
UNIT:POW DBM
CALC:DATA:TPOW? Returns -3.420000E+000
Notes Special NAN (not a number) value
-2251799813685248 is returned if analysis result
could not be computed.
See Also :CALCulate[1..n][:WDM]:DATA:OSNR:FLATness?
:CALCulate[1..n][:WDM]:DATA:OSNR:MEAN?
:CALCulate[1..n][:WDM]:DATA:SIGnalPower:FLA
Tness?
:CALCulate[1..n][:WDM]:DATA:SIGnalPower:ME
AN?
Zobrazit stránku 507
1 2 ... 503 504 505 506 507 508 509 510 511 512 513 ... 637 638