SCPI Command Reference
Optical Spectrum Analyzer 465
Product-Specific Commands—Description
:CALCulate[1..n]:PEAK[:DETEction]:
LEVel
Description This command sets level in relation to peak for
WDM Analysis test.
For this command to be accepted, the OSA
module must be in Ready state with no
acquisition in progress.
At *RST, this value is -45.00 dBm.
Syntax :CALCulate[1..n]:PEAK[:DETEction]:LEVel
<wsp><PeakDetectionLevel>
Parameter(s) PeakDetectionLevel:
The program data syntax for
<PeakDetectionLevel> is defined as a
<DECIMAL NUMERIC PROGRAM DATA>
element.
Selects the Peak detection level.
Range: [-99.99 … 10.00] dBm
Example(s) CALC:PEAK:DETEction:LEVel -65.08
See Also CALCulate[1..n]:PEAK:DETEction:LEVel?