SCPI Command Reference
498 FTB-5240S/5240BP
Product-Specific Commands—Description
:SENSe[1..n]:AVERage:INBand:COUNt
Description This command sets the number of In-Band scans
that will be averaged to produce final trace.
For this command to be accepted, the OSA
module must be in Ready state with no
acquisition in progress.
At *RST, this value is 64.
Syntax :SENSe[1..n]:AVERage:INBand:COUNt<wsp>
<InbandCount>
Parameter(s) InbandCount:
The program data syntax for <InbandCount> is
defined as a <numeric_value> element.
Selects the Number of scans used to inband for
the final scan.
Range: [4 ... 9999]
Example(s) SENS:AVER:TYPE AVERAGING
SENS:AVER:INBand:COUN 4
SENS:AVER:COUN 12
INITiate:IMMediate
See Also SENSe[1..n]:AVERage:INBand:COUNt?
SENSe[1..n]:AVERage:TYPE
Zobrazit stránku 505