Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Uživatelský manuál Strana 504

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 537
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 503
SCPI Command Reference
496 FTB-5240S/5240BP
Product-Specific Commands—Description
:SENSe[1..n]:AVERage:COUNt
Description Sets the number of scans that will be averaged to
produce final trace.
For this command to be accepted, the OSA
module must be in Ready state with no
acquisition in progress.
At *RST, this value is 8.
Syntax :SENSe[1..n]:AVERage:COUNt<wsp><AverageC
ount>
Parameter(s) AverageCount:
The program data syntax for <AverageCount> is
defined as a <numeric_value> element.
Selects the number of scans that will be
averaged to produce final trace.
Range: [2 ... 9999]
Example(s) SENS:AVER:TYPE AVERAGING
SENS:AVER:COUN 12
INITiate:IMMediate
See Also SENSe[1..n]:AVERage:COUNt?
SENSe[1..n]:AVERage:TYPE
Zobrazit stránku 503
1 2 ... 499 500 501 502 503 504 505 506 507 508 509 ... 536 537

Komentáře k této Příručce

Žádné komentáře