Analyzing Traces and Events
OTDR 169
Setting the Analysis Detection Thresholds
To set the analysis detection thresholds:
1. From the main window, press Setup.
2. From the Setup dialog box, select the Acquisition tab.
3. Under Detection Threshold Analysis, set the parameters.
³ Enter the desired values in the appropriate boxes.
OR
³ Under Detection Threshold Analysis, select the default settings by
pressing Default.
4. Press Apply to confirm the changes, then OK to return to the main
window.
The analysis detection thresholds you have just set are applied to all newly
acquired traces. It is also possible to change these thresholds for a specific
trace for reanalysis. For details, see Viewing and Modifying Current Trace
Settings on page 153.