Exfo FTB-7000 OTDR Series for FTB-500 Uživatelský manuál Strana 380

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 482
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 379
SCPI Command Reference
366 OTDR
Product-Specific Commands—Description
:CONFigure[1..n]:ACQuisition:
DURation
Description This command specifies the duration that will be
used for the next acquisition.
*RST reverts this setting to default value.
Syntax :CONFigure[1..n]:ACQuisition:DURation<wsp>
<Duration>|MAXimum|MINimum|DEFault
Parameter(s) Duration:
The program data syntax for <Duration> is
defined as a <numeric_value> element. The
<Duration> special forms MINimum, MAXimum
and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
DEFault allows the instrument to select a value
for the <Duration> parameter.
Sets the acquisition duration, in seconds.
Example(s) CONF:ACQ:DUR? Ex.: Returns 15
CONF:ACQ:DUR 10
CONF:ACQ:DUR? Returns 10
See Also FETCh[1..n]:DURation?
FETCh[1..n]:ASETting:DURation?
Zobrazit stránku 379
1 2 ... 375 376 377 378 379 380 381 382 383 384 385 ... 481 482

Komentáře k této Příručce

Žádné komentáře