SCPI Command Reference
OTDR 353
Product-Specific Commands—Description
Parameter(s) ³ Label:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
TRC1|TRC2|TRC3|TRC4.
Trace index of the available wavelengths.
³ End-of-Fiber:
The program data syntax for <End-of-Fiber> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
Sets the end-of-fiber threshold.
Example(s) CONF:ANA:THR:EOF 5.1
CONF:ACQ:MODE ACQ
INIT
INIT:STAT? Returns 0 when acquisition is
complete.
CALC:THR:EOF? TRC1 Returns 5.1
CALC:THR:EOF TRC1,5.2
CALC:THR:EOF? TRC1 Returns 5.2
See Also CALCulate[1..n]:ANAlysis:[UNIDirectional]
CALCulate[1..n]:EVENt:COUNt?
CALCulate[1..n]:EVENt?
MMEMory[1..n]:LOAD:TRACe
TRACe[1..n]:CATalog?
:CALCulate[1..n]:THReshold:EOFiber