SCPI Command Reference
350 OTDR
Product-Specific Commands—Description
Parameter(s) ³ MarkerB:
The program data syntax for <MarkerB> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
Specifies the marker B position, in meters.
³ SubMarkerB:
The program data syntax for <SubMarkerB> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
Specifies the submarker B position, in meters.
Return the splice loss value, calculated using all
four markers.
Response Syntax <Splice Loss>
Response(s) Splice Loss:
The response data syntax for <Splice Loss> is
defined as a <NR3 NUMERIC RESPONSE DATA>
element.
Return the splice loss value, calculated using all
four markers.
Example(s) CONF:ACQ:MODE ACQUISITION
INIT
INIT:STAT? Returns 0 when acquisition is
complete.
CALC:SLOS? TRC1,10,100,200,300 Ex.: Returns
0.058
CALC:SLOS? TRC1,0.01 KM,100 M, 0.2 KM,300 Ex.:
Returns 0.058