Exfo FTB-7000 OTDR Series for FTB-500 Uživatelský manuál Strana 451

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 482
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 450
SCPI Command Reference
OTDR 437
Product-Specific Commands—Description
:SOURce[1..n]:VFLocator:AM:INTernal:
FREQuency?
Description This query returns a value indicating the current
internal modulation frequency. If the visual fault
locator (VFL) is in CW mode, the function will
return 0.
*RST sets the modulation frequency to 0 Hz
(CW).
Syntax :SOURce[1..n]:VFLocator:AM:INTernal:FREQuen
cy?[<wsp>MINimum|MAXimum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MINimum|MAXimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Frequency>
Zobrazit stránku 450
1 2 ... 446 447 448 449 450 451 452 453 454 455 456 ... 481 482

Komentáře k této Příručce

Žádné komentáře