Exfo IQS-2600B Tunable Laser Source for IQS Platforms Uživatelský manuál Strana 115

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 142
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 114
SCPI Command Reference
Tunable Laser Source 109
Product-Specific Commands—Description
:SOURce[1..n]:WAVelength:SWEep:
SPEed
Description This command allows you to select the sweep
speed for continuous mode.
At *RST, the sweep speed is 15 nm/s.
Syntax :SOURce[1..n]:WAVelength:SWEep:SPEed<wsp
><Speed>|MAXimum|MINimum|DEFault
Parameter(s) Speed:
The program data syntax for <Speed> is defined
as a <numeric_value> element. The <Speed>
special forms MINimum, MAXimum and DEFault
are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
DEFault allows the instrument to select a value
for the <Speed> parameter.
This parameter sets the optical spectrum
displacement sweep speed in meters/s.
Example(s) SOUR:WAV:SWE:SPE 35.455E-9 (in meters/s)
Notes This function cannot be performed while the
module is sweeping.
See Also SOURce[1..n]:WAVelength:SWEep:SPEed?
Zobrazit stránku 114
1 2 ... 110 111 112 113 114 115 116 117 118 119 120 ... 141 142

Komentáře k této Příručce

Žádné komentáře