SCPI Command Reference
90 IQS-2600B
Product-Specific Commands—Description
:SOURce[1..n]:WAVelength:SWEep:MODE
Description This command allows you to select the sweep
mode.
At *RST, sweep mode is continuous.
Syntax :SOURce[1..n]:WAVelength:SWEep:MODE<wsp
>CONTinuous|STEPped
Parameter(s) SweepMode:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
CONTinuous|STEPped.
This parameter is a value representing the sweep
mode:
CONTinuous selects Continuous sweep mode.
STEPped selects the Step-by-Step sweep mode.
Example(s) SOUR:WAV:SWE:MODE CONT
Notes This function cannot be performed while the
module is sweeping.
See Also SOURce[1..n]:WAVelength:SWEep:MODE?