Exfo IQS-2600B Tunable Laser Source for IQS Platforms Uživatelský manuál Strana 59

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 142
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 58
Maintenance
Tunable Laser Source 53
Adjusting Your Unit According to Wavelength
The difference between the pre-selected wavelength and the measured
wavelength cannot be greater than ± 0.200 nm. If the difference between
the measured and preselected wavelengths is greater than ± 0.200 nm
(for example, if you have entered the wrong value), an error message will
be generated.
IMPORTANT
Note that the offset introduced into the Tunable Laser Source with
this software feature cannot be disabled. To correct for a handling
error during the procedure, you must repeat the steps described
above with a calibrated wavelength meter.
Zobrazit stránku 58
1 2 ... 54 55 56 57 58 59 60 61 62 63 64 ... 141 142

Komentáře k této Příručce

Žádné komentáře