SCPI Command Reference
Tunable Laser Source 89
Product-Specific Commands—Description
:SOURce[1..n]:WAVelength:MODE?
Description This query returns the current source mode.
At *RST, resolution mode is NORMal.
Syntax :SOURce[1..n]:WAVelength:MODE?
Parameter(s) None
Response Syntax <SourceMode>
Response(s) SourceMode:
The response data syntax for <SourceMode> is
defined as a <CHARACTER RESPONSE DATA>
element.
The <SourceMode> response is the current
source resolution mode.
NORMal, the source is in Normal operation
mode.
HRESolution, the source is in High Resolution
operation mode.
Example(s) SOUR:WAV:MODE HRES
SOUR:WAV:MODE? returns HRESOLUTION (the
source is in High Resolution operation mode)
See Also SOURce[1..n]:WAVelength:MODE
Zobrazit stránku 94