Zobrazit stránku 378
SCPI-Befehlsreferenz
OSA 371
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
CENTer:ITUGrid
Description This command controls activation of snap center
on ITU grid for WDM analysis default channel.
At *RST, snap center on ITU grid is set to off
(disabled).
Syntax :CALCulate[1..n][:WDM]:CHANnel:AUTO:CENTer
:ITUGrid<wsp><Auto>
Parameter(s) Auto:
The program data syntax for <Auto> is defined
as a <Boolean Program Data> element. The
<Auto> special forms ON and OFF are accepted
on input for increased readability. ON
corresponds to 1 and OFF corresponds to 0.
The <Auto> parameter corresponds to the new
state of the snap center on ITU grid.
0 or OFF: disable default channel snap center on
ITU grid.
1 or ON: enable default channel snap center on
ITU grid.
Snap default channel center on ITU grid enable
state