SCPI-Befehlsreferenz
OSA 539
Product-Specific Commands—Description
SENS:WAV:STAR 1525.000 NM
SENS:WAV:STOP 1570.000 NM
TRIG:SEQ:SOUR IMM
STAT? Poll until returned state is READY
INIT:CONT ON
INIT:CONT? Returns 1 (trigger system
continuously initiated)
...
INIT:CONT OFF
STAT:OPER:BIT8:COND? Poll until returned state
is 0
Notes The trigger system leaves IDLE state to perform
acquisition only if the instrument is in READY
status.
Trace averaging is not supported by the trigger
system when continuously initiated.
Continuous acquisition does not support InBand
noise analysis: acquired trace is always analysed
using IEC noise measurement.
See Also :ABORt
:CALCulate[1..n][:WDM]:STATe
:INITiate[:IMMediate]
:INITiate:CONTinuous?
:SENSe[1..n]:AVERage:STATe
:STATus?
:STATus:OPERation:BIT<8|9>:CONDition?
:TRACe:FEED:CONTrol
:TRIGger[1..n][:SEQuence]:SOURce
:INITiate:CONTinuous