SCPI-Befehlsreferenz
OSA 537
Product-Specific Commands—Description
:INITiate:CONTinuous
Description This command is used to select whether the
trigger system is continuously initiated or not.
The trigger system is used to control trace
acquisition.
At *RST, this value is set to off (disabled).
Syntax :INITiate:CONTinuous<wsp><Continuous>
Parameter(s) Continuous:
The program data syntax for <Continuous> is
defined as a <Boolean Program Data> element.
The <Continuous> special forms ON and OFF
are accepted on input for increased readability.
ON corresponds to 1 and OFF corresponds to 0.
The <Continuous> parameter corresponds to
the new state of the trigger system continuous
cycle.
0 or OFF: disable continuous cycle: trigger
system returns to idle.
1 or ON: enable continuous cycle.