Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Uživatelský manuál Strana 432

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 653
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 431
SCPI-Befehlsreferenz
424 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:
CENTer[:WAVelength]
Description This command sets the nominal center
wavelength of the selected WDM analysis
channel.
At *RST, this value is not available.
Syntax :CALCulate[1..n][:WDM]:CHANnel:CENTer[:WAV
elength]<wsp><Center[<wsp>M]>|MAXimu
m|MINimum|DEFault
Parameter(s) Center:
The program data syntax for <Center> is
defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> element is M. The <Center>
special forms MINimum, MAXimum and DEFault
are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
Zobrazit stránku 431
1 2 ... 427 428 429 430 431 432 433 434 435 436 437 ... 652 653

Komentáře k této Příručce

Žádné komentáře