Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Uživatelský manuál Strana 474

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 653
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 473
SCPI-Befehlsreferenz
466 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:
SIGnalPower:TYPE
Description This command selects signal power
measurement type for WDM analysis of the
selected channel.
At *RST, this value is not available.
At CALCulate[1..n][:WDM]:CHANnel[:DEFine],
this value is set to IPOWer (integrated power).
Syntax :CALCulate[1..n][:WDM]:CHANnel:SIGnalPower:
TYPE<wsp>IPOWer|PPOWer|TPOWer
Parameter(s) Type:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
IPOWer|PPOWer|TPOWer.
The parameter corresponds to the newly
selected signal power type.
IPOWer: selects integrated signal power type.
PPOWer: selects peak signal power type.
TPOWer: selects channel total power type.
Zobrazit stránku 473
1 2 ... 469 470 471 472 473 474 475 476 477 478 479 ... 652 653

Komentáře k této Příručce

Žádné komentáře