Testing Multiple DUTs with the LabVIEW Application
DWDM Passive Component Test System 135
Adjusting Result Settings
To set up an Analysis graph:
1. Select the DUT, the In port and the Out port for whose results you want
to view in a graph.
2. Select Analysis.
3. Select the desired curve type from the provided list.
4. Confirm by clicking Apply.
To set up an Acquisition graph:
1. Select the DUT, the In port and the Out port whose results you want to
view in a graph.
2. Select Acquisition.
3. Select the desired trace type (IL, PDL, ORL).
4. Specify the sequence results you want to see by indicating the values
in the Sequence From and Sequence To boxes.
5. Confirm by clicking Apply.