SCPI Commands Reference
DWDM Passive Component Test System 233
Specific Commands
Specific Commands
INITialization:INITialize
Description This command performs an initialization of all IQS modules.
Syntax INIT:INIT
Parameter None
Response None
Example INIT:INIT
Note When INIT:SIMU = 1, the object simulator is used to develop
without IQS modules.
This is a blocking function i.e., In progress appears in the
status bar and you must wait before sending a new command.
INITialization:INITialize?
Description The response to this query indicates if an initialization of the
modules has been performed.
Syntax INIT:INIT?
Parameter None
Response A boolean response representing the state of initialization.
“1” - The system is initialized.
“0” - The system is not initialized.
Example INIT:INIT?