Exfo IQS-12004B DWDM Passive Component Test System for Uživatelský manuál Strana 168

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 437
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 167
Using the Power Meter Linearity Verification Tool
156 IQS-12004B
Nulling Electrical Offsets
To perform a nulling:
1. Put protective caps on power meter detectors as described in the
Instructions zone.
2. Click Nulling to start the null measurement. The nulling is performed
on all power meters at the same time.
When the operation is complete, the Start button becomes enabled
and the Instructions zone indicates that the tool is ready for linearity
verification.
Note: Once the hardware initialization is complete, you can perform a nulling at
any moment, except during an acquisition.
Zobrazit stránku 167
1 2 ... 163 164 165 166 167 168 169 170 171 172 173 ... 436 437

Komentáře k této Příručce

Žádné komentáře