SCPI Commands Reference
DWDM Passive Component Test System 241
Specific Commands
CONFiguration:STARt
Description This command sets the position where to start the acquisition.
Syntax CONF:STAR<wsp><wavelength>
Parameter The <wavelength> parameter value indicates the wavelength
using the 9999.999 nm format.
Response None
Example CONF:STAR 1520.500
CONFiguration:STARt?
Description The response to this query indicates the position where to start
the acquisition.
Syntax CONF:STAR?
Parameter None
Response Wavelength in 9999.999 nm format.
Example CONF:STAR?