COM Objects Reference
DWDM Passive Component Test System 279
Measures Interface—Methods
Measures
RefIlPdl
Description This command performs an IL/PDL reference measurement.
Syntax object.RefIlPdl(FireEvent)
Parameters Byval Boolean FireEvent
Response None
Possible error ErrInitBefore, ErrCalWrBefore, InProgress, ErrAcqType,
ErrAcqStopped, ErrScanRange, ErrOptimizeBefore,
ErrSourceNotEnabled. For more information, see IQS-12004B
DWDM Passive Component Test System (Standard) COM Error
Messages on page 208.
Note Use only the IQS-5150 polarization state adjuster.
Measures
RefOrlMandrel
Description This command performs an ORL reference measurement by
mandreling the launch fiber.
Syntax object.RefOrlMandrel(FireEvent)
Parameters Byval Boolean FireEvent
Response None
Possible error ErrInitBefore, ErrCalOrlRtjBefore, ErrCalOrlMandrelRtjBefore,
ErrInProgress, ErrAcqType, ErrAcqStopped, ErrScanRange,
ErrSourceNotEnabled. For more information, see IQS-12004B
DWDM Passive Component Test System (Standard) COM Error
Messages on page 208.